4 - 10 MP6000 INTEGRATOR GUIDE
Over Capacity Test
Over Capacity 15.100 EEEE N/A N/A
Decreasing Load Test
15.000 15.000 .0075 kg .0075 kg
10.000 10.000 .005 kg .005 kg
5.000 5.000 .005 kg .003 kg
2.500 2.500 .0025 kg .002 kg
0.100 0.100 .0025 kg .001 kg
Return to Zero Test
0.000 0.000 0.0 kg 0.0 kg
Shift Test (see
Figure 4-4)
5.00 Position 1 5.000 .005 kg .003 kg
5.00 Position 2 5.000 .005 kg .003 kg
5.00 Position 3 5.000 .005 kg .003 kg
5.00 Position. 4 5.000 .005 kg .003 kg
Table 4-5
Verification Tests for Metric Scale (Continued)
Verification Test for
Metric Scale Applied
Load kg
Indication = Applied
Load Within
Applicable
Tolerances
All Tolerances
Are +/-
All Tolerances
Are +/-
Acceptance
Single Int
Acceptance
Dual Int